Specifications

IQmark Digital ECG Operation Manual
25
The live ECG screen displays the results of the 12-channel ECG in the three-by-four (3x4)
leads display format. This format can be changed by using the Settings function. Your
facility’s name, the date, time and the patient’s heart rate appear at the top of the
screen. The ECG tracings are in the center of the screen moving from left to right.
Below the moving tracings are the sweep speed, gain, and filter settings. You can click
the appropriate button to modify its setting. Changes to these settings made from this
screen are temporary and only apply for as long as you are in this test. You can change
and set them as the default settings for all new ECG tests by clicking Settings.
Clicking on INST resets the ECG module and stabilizes the ECG baselines. If there is no
INST button on the screen, manually turning the ECG module power off for one second
and then turning it on would accomplish the same effect.
In the lower-right corner of the screen are the status indicators for ECG Report and RR
Variability. An empty status indicator means that you have not acquired the specific
report since entering the screen. Once a report is created, the indicator next to the test
type appears solid. This indicator is not affected by tests that were previously stored or
reviewed in the database.
The menu bar at the bottom of this screen provides different options you can take.
These available options change according to the settings and the process you have
made.
Print
Prints the last 10 seconds of ECG data being collected. This report contains no
measurements or diagnostic statements. If data is not actively being collected,
this function prints out the last 10 seconds of data in memory.
NOTE: Clicking Print more than once before the report is printed produces
multiple printouts. Exit is disabled until the report is printed. To
speed up printing, click Freeze before clicking Print.
Analyze
Performs resting analysis on the last 10 seconds of data and stores this
analysis as an ECG Report, which can be printed from the Report Review
screen or by selecting Print After Analysis in the Printer Format Settings.
Freeze
Freezes the display and copies the last 10 seconds of ECG data into a memory
buffer. If there are less than 10 seconds of ECG data collected, the freeze
function is delayed until the 10-second buffer is filled. Click this button a
second time to unfreeze the display and clear the memory buffer. When the
display is frozen, you can also click Print or Analyze.
Display
This option is only available if the Display Real-time ECG Tracing setting is
disabled. Click on this button to display the ECG tracing acquired in the last 10
seconds. Tracings are displayed in static form and are not stored. Click on the
Analyze button to analyze and save the ECG displayed.
Go To Table Of Contents