Specifications

IQmark Digital ECG Operation Manual
15
Disable Interpretation
On/Off
Default is Off (cleared).
If On (checked), the ECG will not produce any
diagnostic statements and the interpretation portion
of the report, ECG Review and Edit screen will be
blank.
Disable Quick Edit
On/Off
Default is Off (cleared)
If On (checked), the software defaults to free text
entry method for adding any diagnostic statements,
instead of the Quick Edit feature of the Add a
Statement dialog box. See Editing Diagnostic
Statements.
NOTE: For all pacemaker patients, it is strongly recommended that the ECG be
performed with all filters turned off, particularly the MYO filter. Any artifacts
in the ECG should be corrected at the source, i.e., making sure the electrode
sites are clean of lotion or body hair, the electrodes are fresh and sticky and
are adhering properly on the skin. Refer to ECG Signal Quality Problems in
the Troubleshooting Guide for more details.