INSTALLING THE BATTERY
1) Align the Battery Pack guides with the bottom of the monitor.
2) Slowly close the battery door to ensure the connector in the monitor and the connector
on the battery pack mate together.
3) Lock the battery door closed.
Refer to Page 72, BATTERY POWER for additional battery information.
NOTE:
When the battery pack is re-installed, the monitor will automatically turn “ON”.
WARNING:
Do not disassemble the battery pack or batteries. The batteries contain electrolytes, which can
cause injury to eyes, skin and clothing.
NOTE:
This product contains a rechargeable battery that is recyclable. Under various state and local
laws, it may be illegal to dispose of this battery into the municipal waste stream. Check with your
local authorities for instructions on recycling options in your area.
CHANGING THE FUSES
The CARDELL Monitor uses a dual fuse power input receptacle. The receptacle incorporates
fuses in the hot and neutral AC input lines that are user serviceable.
Refer to Section 11, ACCESSORIES for part number information.
NOTE:
Refer to the monitor’s rear panel label for the proper fuse rating.
CAUTION:
For continued protection against fire hazard, replace only with identically rated fuses.
A fuse may need to be replaced if the monitor is plugged into an electrical outlet but the Battery
Power Visual Indicator is not illuminated the color Green.
WARNING:
Before changing the fuse, unplug the power cord.
The fuse holder is incorporated into the power input receptacle and located under the power
cord input connector.