User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
95
5) The Message Window will display “Chk Pres”, will inflate to approximately 180 mmHg
and attempt to hold this pressure. The pressure value will be displayed in the SYSTOLIC
display window. This test takes about fifteen (15) seconds.
6) At the completion of a successful Pressure Check, the Message Window will display
“Passed”, the monitor will beep two (2) times and will return to the Calibration Check
function after five (5) seconds.
7) If the monitor fails the Pneumatic Pressure Check, the Message Window will display
"Leak", the monitor will beep three (3) times and the return to the Calibration Check
function after five (5) seconds.
8) If the monitor fails the Pressure Check, the Message Window will display "Leak", the
monitor will beep three (3) times and the return to the Calibration Check function after
five (5) seconds.
9) Due to the volume differences of the hoses used with the CARDELL Monitor, the monitor
may incorrectly fail the Plug Tube check. Should the monitor fail the Plug Tube Pressure
Check, obtain a 500 ml Pressure Cylinder and follow the 500 ml Pressure Check.
500 ml PRESSURE CHECK
1) Obtain a fixed volume 500 ml Pressure Cylinder (CAS p/n 01-02-0248).
2) Place the end of the monitor’s inflation hose securely onto the luer fitting at the top of the
pressure cylinder. The hose must fit securely onto the connector for this test to be
performed properly.
3) Enter the Test Mode. The Message Window will briefly display “TestMode” followed by
"0 mmHg".
4) Press the START pushbutton to begin the Pressure Check.
5) The Message Window will display “Chk Prs”, will inflate to approximately 160 mmHg and
attempt to hold this pressure. The pressure value will be displayed in the SYSTOLIC
display window. This test takes about fifteen (15) seconds.
6) At the completion of a successful Pressure Check, the Message Window will display
“Passed”, the monitor will beep two (2) times and will return to the Calibration Check
function after five (5) seconds.
7) If the monitor fails the Pressure Check, the Message Window will display "Leak", the
monitor will beep three (3) times and the return to the Calibration Check function after
five (5) seconds.
8) Should the monitor fail the 500 ml Pressure Check, it is recommended the monitor be
returned to CAS Medical Systems for service.