User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
94
6) Use the manometer inflation bulb to slowly inflate the system pausing for 30 seconds at
the following points and verify calibration according to the following table:
0 mmHg +/- 1 mmHg
50 mmHg +/- 4 mmHg
100 mmHg +/- 4 mmHg
150 mmHg +/- 4 mmHg
200 mmHg +/- 5 mmHg
NOTE:
If the monitor does not display the test pressure for the 30-second period, deflate to zero and
verify the proper assembly of the calibration set-up. Re-inflate the system. If the monitor again
fails to hold the pressure, it is recommended the monitor be returned to CAS Medical Systems
for service.
OVERPRESSURE
1) While still in the Manometer Mode test, inflate the pressure slowly until 290 mmHg +/- 10
mmHg is reached. The Message Window should stop updating, display the message
"OverPres" and provide a NIBP Application Error audible tone.
2) Press the CANCEL pushbutton to exit the Overpressure Test. The monitor returns to the
Manometer Mode selection.
3) If the monitor does not meet the above specifications, it is recommended the monitor be
returned to CAS Medical Systems for service.
PNEUMATIC PRESSURE CHECKS
Check the monitor's pneumatic system for air leakage every six (6) months.
PLUG TUBE
1) Obtain the male luer plug found in the Calibration Kit (product #P9) supplied with the
monitor.
2) Place this plug into the cuff connector at the end of the monitor tubing and twist one-
quarter turn. The plug must fit securely into the connector for this test to be performed
properly.
3) Enter the Test Mode. The Message Window will briefly display “TestMode” followed by
"0 mmHg".
4) Press the START pushbutton to begin the Pressure Check.