User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
93
LED CHECK
The CARDELL Monitor incorporates an all lights “ON” check to verify the functionality of the front
panel LED displays and indicators.
Enter the Test Mode. The Message Window will briefly display “TestMode” followed by
"0 mmHg".
Press and hold either the ARROW UP or ARROW DOWN pushbuttons. The monitor will
illuminate all appropriate 7-segment displays, bar graph, bell icon and patient mode indicators
for as long as the button is pressed.
NIBP CALIBRATION CHECK
Verify the calibration of the monitor once per year.
A Calibration Kit, (product #P9) is included with the monitor. The kit contains a T-connector with
a male and a female luer fitting (for a Calibration Check) and a male luer plug (to be used for the
Pneumatic Check).
Manometer Mode
Overpressure Test
Pneumatic Test
The following equipment is required to perform the pressure checks:
Assembled P9 Calibration Kit
Mercury manometer whose accuracy meets the AAMI/ANSI Standard for Non-Automated
Sphygmomanometers, 2002.
MANOMETER MODE
Assemble the Calibration Kit according to the diagram provided in the P9 kit.
1) Assemble the Calibration Kit according to the diagram provided in the P9 kit.
2) Remove the manometer tubing from the inflation bulb. Connect the open ended tubing
of the T-connector to the inflation bulb.
3) Connect the female luer fitting to the inflation tube leading to the manometer.
4) Connect the male luer fitting to the manometer tubing.
5) Enter the Test Mode. The Message Window will briefly display “TestMode” followed by
"0 mmHg".
Next
Back
Go To Table Of Contents