User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
92
TEST MODE
WARNING:
Do not place the monitor in the TEST MODE when a cuff is attached to a patient.
NOTE:
While in the Test Mode if no pushbutton is pressed within 15 minutes, the monitor will
automatically terminate the Monitor Configuration menu and return to the “Ready” mode.
The monitor must be in the Configuration Mode in order to perform the following functions:
All LEDs “ON” Check
NIBP Calibration Check
o Manometer Mode
o Over Pressure
Pneumatic Pressure Checks
To enter the monitor’s Configuration Menu, press and hold the ALARM LIMITS
and
VOLUME
pushbutton keys while the monitor is being turned “ON” .
Once in the menu, press one of the PREVIOUS (CYCLE TIME / ALARM LIMITS) programmed
pushbutton keys until the Message Window briefly displays “TestMode” followed by “0 mmHg”.
EXIT THE TEST MODE
When you have completed with the Test Mode, press the CANCEL pushbutton to exit. The
Message Window will briefly display “Saving” and return to the “Ready” state.
Next
Back
Go To Table Of Contents