User`s manual

CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
89
PRINTER
When the printer becomes dirty, wipe with a soft dry cloth. For extreme dirt buildup, soak a cloth
with mild detergent, wring well and wipe. Dry by wiping with a soft dry cloth.
CAUTION:
Before cleaning the printer, disconnect the AC adapter from the printer.
Do not use volatile chemicals such as thinner, benzine, etc.
Never wet the inside of the printer mechanism.
Refer to the printer User’s Manual for more information.
SPO
2
INTERCONNECT CABLE
Prior to each patient use, inspect the SpO
2
Interconnect cable for damage. As necessary clean
the cable using a soft cloth dampened with a germicidal solution.
SENSOR AND CLIPS
CAUTION:
To avoid damage to the VetSat sensor, remove it from the clip before cleaning either piece.
CAUTION:
Do not sterilize the sensor or clips by irradiation, steam or ethylene oxide. Do not immerse the
sensors in water or cleaning solution.
When necessary, the sensor may be surface-cleaned by wiping it with an agent such as 70%
Isopropyl Alcohol.
The clip may be cleaned by either wiping it with, or soaking it for ten (10) minutes in, 70%
Isopropyl Alcohol. If the clip is soaked, be sure to rinse it with water and air-dry it prior to use on
the next patient.
After each cleaning and prior to each use, inspect the sensor and cable for fraying, cracking,
breakage, or other damage. Inspect the clip for cracking or breakage, or loss of spring tension
that would allow slippage or movement of the sensor from its proper position.
NOTE:
If defects are noted, do not use the sensor or clip.
Next
Back
Go To Table Of Contents