9. CLEANING
CLEANING OVERVIEW
WARNING:
Do not, under any circumstances, perform any testing or maintenance on the monitor while the
monitor is being used to monitor a patient. The monitor must be turned “OFF”. Unplug the
monitor from the AC power source and remove the internal battery.
CAUTION:
Do not open the monitor to clean or repair it. Contact CAS Medical System for service needs.
CAUTION:
Disconnect all accessories from the monitor before cleaning. Do not immerse any part of the
electrical connector of the cable or accessories in the cleaning or disinfection solution at any
time. Do not use an abrasive cloth or cleaner on the accessories.
THE MONITOR
On a daily basis, examine the monitor case for any damages and check the AC power cord for
bent or broken prongs, cracks or fraying. Neither the monitor nor the power cord should be used
if damaged. If any damage is noted, contact the appropriate service personnel.
CAUTION:
Do not spray any water or cleaning solution directly onto the monitor.
As needed, clean the monitor using a soft cloth dampened with a mild dishwashing detergent
solution and gently rub the soiled area until clean. Use a clean soft cloth to dry the monitor. Do
not use abrasive cleaners on the monitor. Do not use either isopropyl alcohol or solvent to clean
the monitor. Use of these cleaners can cause damage to the monitors’ surface. Do not
immerse the monitor or power cord in the cleaning solution.
When necessary, the monitor surfaces may be disinfected using a soft cloth saturated with a
10% (1:10) solution of chlorine bleach in tap water. When all of the surfaces have been
disinfected, wipe the entire surface of the monitor using a soft cloth dampened with fresh water
to remove any trace amounts of residue and/or fumes.
NOTE:
Thoroughly wipe off any excess cleaning solutions. Care should be taken to prevent water or
cleaning solution to run into connector openings or crevices.