REPLACING THE BATTERY PACK
WARNING:
Do not operate the printer or connect the printer to the CARDELL Monitor with the battery pack
removed.
WARNING:
Never change the battery pack while the battery charger is plugged in and/or the CARDELL
Monitor is being operated.
1) Switch the printer “OFF”.
2) Disconnect the printer from the CARDELL Monitor and unplug the wall charger cord
3) Open the battery door by pressing in on the battery cover and pushing upward.
Figure 14: Opening the Battery Door
4) Remove the battery cover.
5) Remove the battery pack from the compartment and disconnect its connecting cable.