ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)
Table 7: Error Messages on the Message Window
ERROR
MESSAGE
POSSIBLE CAUSE POSSIBLE SOLUTION
"NIBP Err" An electronic failure has
occurred within the NIBP
module.
Contact CAS Medical Systems, Inc. to have
the monitor serviced.
“No Probe” The monitor is not detecting the
SpO
2
probe.
The probe was disconnected.
"OverPres" Cuff pressure exceeded 290
mmHg.
Very rapid squeezing of the cuff can cause
this error.
Repeat the measurement.
If this message repeatedly occurs during
normal use, the monitor must be serviced.
“P Search” The monitor is searching for a
Pulse signal.
Normal at power-up as the monitor searches
for a pulse.
The probe position may have changed.
Check the probe site.
“Prb OFF” The monitor is no longer
receiving a patient signal from
the probe.
The probe is no longer in contact with the
patient.
Check the probe site.
"Pwr Fail" Power was disconnected from
the monitor.
Press the SILENCE/RESET pushbutton to
clear the message.
OR
Re-cycle the monitor’s power.
"RangeErr" The systolic reading exceeds
the measurement range of 265
mmHg.
Repeat measurement.
If the message is displayed again, use
another method to measure the patient’s
blood pressure.
“SetClock” The monitor’s clock needs to be
set.
The monitor’s time and date values are
incorrect.
Refer to Page 70 and for information to set
the Time and Date.
The monitor’s internal clock battery needs to
be replaced. Contact CAS Medical Systems.