ERROR MESSAGES ON THE MESSAGE WINDOW
Table 7: Error Messages on the Message Window
ERROR
MESSAGE
POSSIBLE CAUSE POSSIBLE SOLUTION
"Air Leak" Cuff applied too loosely.
Leak in cuff/monitor
pneumatic system.
Check cuff for snug fit on patient.
Check that the cuff/monitor connection is
secure.
Check cuff for leaks. Do not use a known
leaky cuff.
“Appl Err“ Small Cuff is detected in Large
Cuff Mode.
Check cuff. Replace cuff or change
operating mode.
“ChksumEr“ An electronic failure has
occurred within the monitor’s
Main Control Board.
Contact CAS Medical Systems, Inc. to have
the monitor serviced.
"Dead Bat" The battery is fully discharged. Recharge the battery for at least 4 hours.
"Flow Err" Stable cuff pressure cannot be
maintained by the pneumatic
system.
Check the external tube for kinks.
Perform a Pneumatic Check as detailed in
the Maintenance section of this manual.
Replace cuff.
"LooseCuf" Cuff applied too loosely. Check cuff for proper fit on patient.
"Low Batt" The battery is almost
discharged.
At least 30 minutes of operation are
available from when the message first
appears.
Recharge the battery as soon as possible.
"Motion" There was too much extremity
motion for the monitor to
accurately complete the
measurement in 120 seconds.
Measurements can be obtained when there
is limited extremity movement, but the
measurement time may be extended.
Measurement time is limited to 120
seconds.
Restrain patient extremity motion.
"NBP Cal" Pressure calibration data
corrupted within NIBP module.
Pressure module needs recalibration.
Contact CAS Medical Systems, Inc. to have
the monitor serviced.