User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
74
For the Auto “OFF” feature to work properly, the following conditions must apply;
Alrm Off is selected in the Configuration Menu.
The CARDELL Monitor is operating on battery power.
No NIBP or SpO
2
readings were taken in the past ten (10) minutes.
No key presses were made in the past ten (10) minutes.
The monitor cannot be in the Automatic NIBP cycle mode.
POWER FAIL
The CARDELL Monitor incorporates a Power Fail feature. Whenever the power is disconnected
from the monitor and the monitor is not allowed to shut down in an orderly fashion, the monitor,
when re-powered alerts the user. The message “Pwr Fail” is displayed in the Message Window
and three (3) audio “beeps” followed by two (2) audio “beeps” are heard every ten (10) seconds.
During this condition, all other pushbuttons are inactive except for the POWER and
SILENCE/RESET.
To clear the Power Fail condition, and return to “Ready”, press the SILENCE/RESET pushbutton
or properly recycle the monitor’s power.
USER MESSAGES
The CARDELL Monitor displays a variety of messages to aid the user in monitor operation. If a
troubleshooting message is displayed during a measurement, follow the actions listed to correct
the situation.
If the monitor does not turn on, or exhibits a flashing display and failure to operate, the battery is
most likely below the Dead Battery point. Connect the monitor to an AC wall outlet and allow it
to charge for four (4) hours.
Refer to Table 7, Error Messages on the Message Window for more information.
If the monitor is in need of repair, it must be referred to the appropriate service personnel.
Service performed by unauthorized personnel could be detrimental to the monitor and will void
the warranty. For service, contact CAS Medical Systems, Inc.