User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
70
MAP VALUE ENABLE / DISABLE
During a blood pressure reading, the user can elect to display or not to display the MAP value.
Should it need to be changed, first enter the Configuration menu. Refer to Page 67, ENTERING
THE CONFIGURATION MENU. Once in the menu, use one of the NEXT programmed
pushbutton keys until the Message Window displays the monitor’s current setting for the MAP
value “MAP On” or “MAP Off”.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.
NOTE:
When “MAP Off” is selected, MAP values are omitted from History Display and Printing as well.
All alarms associated with MAP values are also disabled.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
SET THE SpO
2
ALARM DELAY
The delay time until an alarm is generated for %SpO
2
and Pulse Rate can be configured to be
either zero (0) seconds “Delay 0S” or ten (10) seconds “Delay10S”. The default value used by
the CARDELL Monitor is ten (10) seconds.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
SETTING THE DATE
The CARDELL Monitor’s Date value is set at the factory. Should it need to be changed, enter
the Monitor Configuration menu and use one of the NEXT programmed pushbutton keys until
the Message Window displays the monitor’s date using the following format: “DDMMMYY”.
Where DD = Day of the Month, MMM = Month of the Year (Jan, Feb, etc.) and YY = Last 2 digits
of the year (2002 is displayed as 02). The flashing parameter indicates the parameter that can
be changed.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection. Press one of the
NEXT programmed pushbutton keys to advance to the next parameter to set within the Date
menu.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
Next
Back
Go To Table Of Contents