2-MINUTE AUDIO SILENCE
NOTE:
Enabling the 2-Minute Audio Alarm Off mode can only be accomplished after all active alarm
conditions have been addressed. To clear an alarm, refer to Page 63, CLEARING ALARMS for
more information.
If the 2-Minute Audio Off mode is selected in the Monitor Configuration menu (refer to section
MONITOR CONFIGURATION), pressing the SILENCE/RESET pushbutton will temporarily
deactivate the audio alarms for two (2) minutes. The visual indicator next to the
SILENCE/RESET pushbutton will illuminate continuously and the message “2Min Aud” will be
displayed in the Message Window as a reminder.
NOTE:
If the Message Window is currently displaying a time stamp from a previously taken NIBP
measurement, the “2Min Aud” message will alternately be displayed with the time stamp
message.
When the 2-Minute Audio Off is enabled, the monitor will not alarm for patient related High and
Low alarms, NIBP Application Error and SpO
2
Probe Alarm. The monitor will alarm for Low
Battery and Dead Battery.
To re-activate the audio alarm, press the SILENCE/RESET pushbutton again. The visual
indicator will go out. If the button is not pressed, audible alarms will be re-armed automatically
after two (2) minutes. Refer to Page 68, AUDIO ALARMS (SILENCE/RESET Pushbutton) for
more information.
PERMANENT AUDIO ALARM SILENCE
NOTE:
Enabling the Permanent Audio Alarm Off mode can only be accomplished after all active alarm
conditions have been addressed. To clear an alarm, refer to Page 63, CLEARING ALARMS for
more information.
If the Permanent Audio Off mode is selected in the Monitor Configuration menu (refer to section
MONITOR CONFIGURATION), pressing the SILENCE/RESET pushbutton will deactivate the
audio alarms. The visual indicator next to the SILENCE/RESET pushbutton will flash at a rate of
one (1) second “ON” and one (1) second “OFF” and the message “Perm Aud” will be displayed
in the Message Window as a reminder.
NOTE:
If the Message Window is currently displaying a time stamp from a previously taken NIBP
measurement, the “Perm Aud” message will alternately be displayed with the time stamp
message.