AUDIBLE AND VISUAL INDICATORS
The CARDELL Monitor is capable of producing both an audible and a visual indicator for a
variety of monitor conditions. The following table provides a cross reference for audible and
visual indications.
Table 5: Audible and Visual Indicators
Alarm Condition Priority
Level
Audible Tone Visual Indication
Dead Battery High 3 Beeps followed by 2 Beeps
every 10 seconds
The message is displayed
continuously in the Message
Window. The Battery Power
Visual Indicator is Red.
Patient Alarm Limit
Violations
High 3 Beeps followed by 2 Beeps
every 10 seconds
The associated 7-segment LED
flashes for one second on / one
second off for the parameter limit
in violation.
SpO
2
% Alarm Limit
Violations
The message is displayed for one
second every two to six seconds in
the Message Window *.
Power Fail High 3 Beeps followed by 2 Beeps
every 10 seconds
The message is displayed for one
second every two to four seconds
in the Message Window *.
Low Battery Alarm Medium 3 Beeps every 25 seconds The message is displayed for one
second every two to six seconds in
the Message Window *. The
Battery Power Visual Indicator is
Red.
NIBP Application
Error
Medium 3 Beeps every 25 seconds The message is displayed for one
second every two to six seconds in
the Message Window *.
SpO
2
Probe Alarms Medium 3 Beeps every 25 seconds The message is displayed for one
second every two to six seconds in
the Message Window *.
NIBP Complete Low 1 Beep when blood pressure
is completed
Numerics are updated with each
blood pressure taken.
SpO
2
Pulse Beat Low 1 Beep coincides with each
received SpO
2
pulse rate
Numerics are updated with each
received SpO
2
pulse rate value.
Key Click Low 1 Beep associated with each
button action
None
* - The message interval time will vary based on what monitor conditions are present.
NOTE:
Refer to Table 7, on Page 76 for a listing of messages that may be displayed in the Message
Window.