User`s manual
CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
61
The Message Window will display “Saving” and the current value(s) will be saved. When
finished, press the CANCEL pushbutton to return to the “Ready” state. These values will be
retained in memory even after the monitor is turned “OFF”.
RESTORE FACTORY DEFAULTS
To restore the factory default settings, press ALARM LIMITS pushbutton to enter the Alarm
Limits mode.
Press and Hold the ALARM LIMITS pushbutton for 2 seconds until the message “Saving? N”
appears in the Message Window.
Press the ALARM LIMITS pushbutton to bring up the Recall menu. The message “Recall? N”
appears in the Message Window. Pressing either the ARROW UP or ARROR DOWN
pushbutton will display the message “Recall? Y” in the Message Window.
Press the ALARM LIMITS pushbutton. The Message Window will display "Restore" and return
to the Alarm Limit menu. When finished, press the CANCEL pushbutton to return to the “Ready”
state.
ALARM LIMIT VALUES
Table 4, lists the Alarm Limit Default Values used by the CARDELL Monitor.
Table 4: Default Alarm Values
Systolic Diastolic MAP Pulse %SpO
2
High 240 130 220 290 OFF
Low OFF OFF OFF OFF 88
NOTE:
Patient alarms for Systolic, Diastolic and MAP values are produced at the time the measurement
is taken. Alarms for %SpO
2
have a ten (10) second delay. Alarm for Pulse Rate will be
immediate if taken from the NIBP or have a ten (10) second delay if taken from the SpO
2
signal.