PATIENT ALARM MODE
WARNING:
Configuring the CARDELL Monitor’s alarm settings to “OFF” will disable all audible and visual
alarms. This mode should only be selected for spot check applications where the patient is
receiving bedside surveillance by a trained clinician. Refer to Page 68, AUDIO ALARMS
(SILENCE/RESET Pushbutton) for more information.
The CARDELL Monitor is equipped with patient alarms to warn the user if any measurement
parameter is outside the range of a user set value. This feature will allow the user to set values
for:
Systolic High and Low
Diastolic High and Low
MAP High and Low
Pulse High and Low
%SpO
2
High and Low (Model 9402)
CHANGING ALARM LIMITS
The CARDELL Monitor incorporates a two (2) step process for reviewing and changing Alarm
Limit values.
Press the ALARM LIMITS pushbutton once and the monitor’s front panel displays indicate the
current values of all High Alarm Limit values, for two (2) seconds then toggles to indicate all Low
Alarm Limit values for two (2) seconds. This reviewing of all High and Low Alarm Limit values
continues for thirty (30) seconds if no other pushbutton is pressed.
To change the value of an Alarm Limit, press the ALARM LIMITS pushbutton again. The Alarm
Limit value is displayed in its corresponding display window and the Message Window will
display the limit’s name.
Use the ARROW UP or ARROW DOWN pushbuttons to change the limit value. Use the
ALARM LIMITS pushbutton key to advance onto the next parameter value.
Press the CANCEL pushbutton to return to "Ready" or after thirty (30) seconds of button
inactivity the monitor will automatically return to "Ready". The alarm value(s) set will now be
used until power is turned “OFF”.
SAVING ALARM LIMITS
To save the Alarm Limit values to non-volatile memory, press and hold the ALARM LIMITS
pushbutton for two (2) seconds while in the Alarm Limits set value mode. The Message Window
will display "Save? N".
Press either the ARROW UP or ARROW DOWN pushbuttons will display “Save? Y” in the
Message Window. Press the ALARM LIMIT pushbutton a final time.