PRINT HISTORY (to optional printer)
The stored Event History and Trend History data are printed individually. Press and Hold the
HISTORY pushbutton for two (2) seconds while in either the Event or Trend History screens.
The message “Print? N” appears in the Message Window.
Use either the ARROW UP or ARROW DOWN pushbutton until the message “Print? Y” appears
in the Message Window.
Press the HISTORY pushbutton again to start printing the History data. The message “Printing”
appears in the Message Window as a status indicator.
NOTE:
Prior to accessing the Print History mode, the printer should have paper installed and the power
turned “ON”. Refer to Section 8 for more information about printing and sample printouts of both
History screens.
CLEARING HISTORY
The History data can be manually cleared by pressing and holding for two (2) seconds the
HISTORY pushbutton while in either of the History modes. The message “Print? N” appears in
the Message Window.
Press the HISTORY pushbutton again and the message “Erase? N” appears in the Message
Window.
Use either the ARROW UP or ARROW DOWN pushbutton until the message “Erase? Y”
appears in the Message Window.
Pressing the HISTORY pushbutton a final time and the message “Erasing” appears in the
Message Window. The monitor erases the memory, exits the History mode and returns to the
“Ready” state.
REAL TIME CLOCK (RTC)
The CARDELL Monitor uses an internal Real Time Clock to time stamp all entries that are stored
in History. Changes made to either the time or date settings should be performed in-between
patients being monitored. Refer to Page 53, DISPLAYING THE TIME and Page 71, SETTING
THE TIME for more information.