HISTORY MODES
The History Mode allows the user to recall previously taken measurements. The measurement
results and the time the measurement was taken are also displayed.
Patient History is organized in two visually displayed lists.
Event History:
Pressing the HISTORY pushbutton the first time reveals the first viewable list that
contains up to 480 entries of NIBP Event History readings along with the corresponding %SpO
2
results (Model 9402).
Trend History:
Pressing the HISTORY pushbutton again reveals the second viewable list that
contains up to 480 entries of Trend History at one (1) minute intervals. The Trend History
contains all the events from the Event History log combined with a one (1) minute average of the
SpO
2
values (Model 9402) that are saved every minute.
NOTE:
Both Event and Trend History data are only available for twenty-four (24) hours. Data older than
twenty-four (24) hours is automatically removed from the History lists.
Once the HISTORY pushbutton has been pressed, use the HISTORY pushbutton to toggle
between the two History menus.
When completed, press the CANCEL pushbutton to return to "Ready" or after thirty (30) seconds
of button inactivity the monitor will automatically return to "Ready".
NOTE:
History data values for NIBP are saved as they occur. History data values for %SpO
2
are saved
once a minute (one (1) minute average).
NOTE:
Turning the power “OFF” does not clear the History memory. Measurements will remain in
memory for up to twenty-four (24) hours. Any measurement older than twenty-four hours is
deleted. It is suggested to manually clear History between patients (See CLEARING HISTORY