Press the CANCEL pushbutton, at any time, to stop a measurement and deflate the cuff during
the measurement process. The monitor’s front panel NIBP displays will indicate all dashes “- -“
and the Message Window returns to “Ready”.
The CANCEL pushbutton can be pressed after a measurement is taken to clear the current
reading from the displays. The monitor’s front panel NIBP displays will indicate all dashes “- -“
and the Message Window returns to “Ready”.
NOTE:
If any displayed NIBP measurement were to be left on the display for up to twenty-four (24)
hours, the monitor will automatically blank the displays to all dashes “- -“ and the Message
Window returns to “Ready”.
AUTOMATIC CYCLE FOR BLOOD PRESSURE DETERMINATION
The CARDELL Monitor can automatically take blood pressure measurements at pre-selected
time intervals.
To choose a time interval, press the CYCLE TIME pushbutton. The Message Window displays
the format “Auto XX”. Where “XX” is the time interval in minutes.
Choose the time interval using one of the following methods:
Press the CYCLE TIME pushbutton to advance through the available time intervals.
Press the ARROW UP or ARROW DOWN pushbutton keys to advance forward or backwards
through the available time intervals.
Once a time interval has been selected, press the START pushbutton to begin the first
measurement.
Between each measurement, the Message Window will display the time remaining until the next
measurement, as well as the cycle time chosen in the format "MM:SS CC" where “MM” is the
minutes, “SS” is the seconds and “CC” is the cycle time selected.
The measurement results are displayed on the front panel until the start of the next
measurement cycle.
NOTE:
If a measurement is desired between measurement cycles, press the START pushbutton. After
this measurement, the monitor will re-enter the Automatic Cycle mode and countdown to the
next measurement based on the Cycle time selected.
In the Automatic Mode, while during a measurement or in the idle time between measurements,
pressing the CANCEL pushbutton will cause the monitor to exit the Automatic mode and return
to the Manual mode. The monitor’s front panel NIBP numeric displays will indicate the reading
of the previous measurement taken and the Message Window will display the time stamp of that
reading.