MONITOR OPERATING INSTRUCTIONS
TURNING THE CARDELL MONITOR “ON”
Press the ON/STANDBY pushbutton on the front panel to turn the monitor “ON”.
Figure 10: Turning the Monitor On
Each time the monitor is turned “ON”, a brief electronic song will play while the Power On Self-
Test (POST) is conducted to ensure that its internal circuits are functioning properly.
NOTE:
The user should use the Power On Self Test as a verification tool that all front panel visual
indicators and the audio are functioning properly.
The one (1) second Configuration Setup Test is a visual indication of the CARDELL Monitor’s
Model number (9401 or 9402).
The four (4) second Power On Self-Test consists of:
The monitor plays its start-up song from the internal speaker.
All equipped parameter segments are lit for one (1) second.
All High Alarm Values are displayed for one (1) second in their corresponding numeric
display window and an audible tone is emitted from the monitor’s internal speaker. The
Message Window indicates “HI LIMS”.
All Low Alarm Values are displayed for one (1) second in their corresponding numeric
display window and an audible tone is emitted from the monitor’s internal speaker. The
Message Window indicates “LO LIMS”.
The current time is displayed for 1 second.