5. BLOOD PRESSURE MONITORING
CUFF SELECTION AND APPLICATION
The use of properly designed and sized cuffs is essential for the accurate non-invasive
measurement of blood pressure. Midmark cuffs are recommended for use with the CARDELL
Monitor.
NOTE:
CAS recommends the use of Midmark cuffs with the CARDELL Monitor.
CAS recommends the use of the inflation hose supplied with the monitor or a replacement from
CAS.
The widest cuff that can be placed around the limb should be used. A cuff that is too small for
the limb will not supply sufficient occlusion pressure to the artery. This can cause an
erroneously high blood pressure reading. Ideally, cuff width should be 40% of the limb
circumference. Substitution of a cuff different from that supplied might result in a measurement
error.
NOTE:
Overlapping the cuff will not affect the measurement results.
For best results, a cuff should be wrapped for a snug fit and be positioned reasonably close to
heart level. Measurements made above the level of the heart will give reduced blood pressure
readings. Measurements made below the heart level will give increased readings. These errors
are mainly due to the weight of the blood.
NOTE:
Avoid compression or restriction of the NIBP Inflation Hose. The hose must not be kinked or
pinched. It can be placed in any position, but it should be kept off the table surface to avoid
equipment vibrations.
WARNING:
The cuff should not be applied on a limb being used for an intravenous infusion. Do not place
the cuff on any extremity being used for SpO
2
monitoring.
See Section 11, ACCESSORIES for CAS Medical Systems cuff size and part number