WARNING:
ACCURACY – If the accuracy of any measurement does not seem reasonable, first check the
patient’s vital signs by alternate means and then check the CARDELL Monitor for proper
functioning.
CABLES – Route all cables away from patient’s throat to avoid possible strangulation.
DEFIBRILLATION – Do not come in contact with patients during defibrillation. Serious injury or
death could result.
DISPOSAL – Dispose of the packaging material, observing the applicable waste control
regulations.
LEAKAGE CURRENT TEST – The interconnection of auxiliary equipment with this device may
increase the total leakage current. When interfacing with other equipment, a test for leakage
current must be performed by a qualified biomedical engineering personnel before using with
patients. Serious injury or death could result if the leakage current exceeds applicable
standards.
SITE REQUIREMENTS – For safety reasons, all connectors for patient cables and sensor leads
are designed to prevent inadvertent disconnection, should someone pull on them. Do not route
cables in a way that they may present a stumbling hazard. For devices installed above the
patient, adequate precautions must be taken to prevent them from dropping on the patient.