User`s manual

CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
107
POWER
Characteristic Specification
Source: External line or internal battery
AC Power Option: 100 - 240 VAC, 50/60 Hz, 0.5A; Fuse Rating – T500mAL250V or
T1.25AL250V (two provided) Refer to monitor rear panel labeling for actual
fuse rating.
Battery: Nickel Metal Hydride (NiMH) battery pack (user removable)
Charge Time: 4 hours
Operation on battery: 100 NIBP readings when set in the 5-minute
Automatic Mode.
Leakage Current: 100 microamp (maximum)
FEATURES
Characteristic Specification
Self Test: System self test is performed each time power is turned on.
Auto Zero: Zero pressure reference is automatically established after every reading.
Inflation: Initial inflation to 150 mmHg or user selectable (100, 120, 140, 150, 160,
180, 200). Subsequent inflation to approximately 30 mmHg greater than
previous Systolic pressure.
Deflation: Automatic
Max Measurement Time: Limited to 150 seconds
SAFETY LIMITS
Characteristic Specification
Automatic Cuff Deflation: If cuff pressure exceeds 290 mmHg
If measurement time exceeds 150 seconds
If safety timer detects microprocessor failure
Next
Back
Go To Table Of Contents