User`s manual

CARDELL
®
Veterinary Monitors
21-02-0285 REV. 06 07/11
105
13. SPECIFICATIONS
NIBP MEASUREMENT
Characteristic Specification
Technique:
Oscillometric (MAXNIBP
® Technology)
Microprocessor software eliminates most ambient noise and motion artifact.
Blood Pressure Range
Systolic: 25 – 265 mmHg
Diastolic: 15 – 220 mmHg
MAP: 20 – 235 mmHg
Pulse Rate Range: 30 – 300 BPM
Accuracy
Blood Pressure: +/-5 mmHg with a standard deviation no greater than 8 mmHg (See
Standards)
Pulse Rate: +/-2% or +/-2 BPM, whichever is greater
%SpO
2
MEASUREMENT
Characteristic Specification
Nellcor®
Type: Functional Oxygen Saturation
SpO
2
% Range: 1 - 100%
SpO
2
Accuracy: 70 - 100%, +/-2 digits (1 S.D.) Adult
70 - 100%, +/-3 digits (1 S.D.) Neonate
Measurement
Wavelengths:
Red 660 Nanometers
Infrared 890 Nanometers
Power: Not exceeding 15 mW
Pulse Rate Range: 20 - 300 BPM
Pulse Rate Accuracy: +/-3 digits
Numerics: Updated every one (1) second.
NOTE:
Accuracy is specified for adult human hemoglobin measured at the fingertip. Although animal
hemoglobin has similar optical characteristics, other types of hemoglobin may affect the
accuracy.