Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
80
REPLACING THE NIBP MODULE
The NIBP Module is part of the Front Case Assembly.
Removal
1) Complete the steps in Prior to Disassembly and Monitor Disassembly.
2) Disconnect the NIBP hose from the NIBP connector located on the side panel. Note the
location of the hose from the connector to the board.
3) Remove the three (3) screws that secure the NIBP Module.
4) Separate the NIBP Module from the J4 connector located on the Main Control board.
Figure 13: Replacing the NIBP Module
Installation
1) Align the female connector on the NIBP Module with the J4 connector on the Main
Control board. Gently push the NIBP Module onto the connector pins.
2) Secure the board to the standoffs using the Loctite adhesive and the hardware
previously removed. Apply a small amount of the adhesive to the threads of the screws
and torque to 30 in.-oz.
3) Re-route the hose under the board and connect it to the NIBP connector on the side
panel.
4) Finish assembling the monitor by following the steps in section Monitor Assembly.
Back
Next