REPLACING THE MONITOR BATTERY
A part number for the battery can be found on the label located on the inside panel of the
battery pack. When the battery fails to hold a charge it will need to be replaced.
CAS Medical Systems recommends the battery be changed every two (2) years.
REMOVING THE BATTERY
1) Turn the monitor “OFF” and disconnect the power cord.
2) Push down on the battery latch to unlock the battery door from the rear panel of the
monitor.
3) Carefully remove the battery pack from the rear panel of the monitor. Refer to Figure 11.
Figure 11: Removing the Battery
INSTALLING THE BATTERY
1) Align the Battery Pack guides with the bottom of the monitor.
2) Slowly close the battery door to ensure the connector in the monitor and the connector
on the battery pack mate together.
3) Lock the battery door closed.
NOTE:
When the battery pack is re-installed, the monitor will automatically turn “ON”.