CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
74
LED Check
Verify, all front panel LED displays and visual indicators are lit. Pass ( ) Fail ( )
+12 Volt Power Supply Check
Verify, the voltage reading to be 12 +/- 0.50. Pass ( ) Fail ( )
System Pressure
Deflate system pressure to 0 mmHg. Pass ( ) Fail ( )
Inflate system pressure to 50 mmHg. Pass ( ) Fail ( )
Inflate system pressure to 100 mmHg. Pass ( ) Fail ( )
Inflate system pressure to 150 mmHg. Pass ( ) Fail ( )
Inflate system pressure to 200 mmHg. Pass ( ) Fail ( )
Over Pressure
Verify, the Message Window displays “Over Pres”. Pass ( ) Fail ( )
Pneumatic Pressure Tests
Plug Tube Test Pass ( ) Fail ( )
500 mL Pressure Check Pass ( ) Fail ( )
NIBP Simulator Check
Monitor displays correct NIBP values. Pass ( ) Fail ( )
Oximetry Simulator Check
Monitor displays correct SpO
2
values. Pass ( ) Fail ( )
Leakage
Verify, the monitor’s leakage to be less than 100 micro-amps. Pass ( ) Fail ( )
Hypot
Monitor, verify, no arching was observed. Pass ( ) Fail ( )
SpO
2
, verify, no arching was observed. Pass ( ) Fail ( )