CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
71
HYPOT (Monitor)
NOTE:
CAS Medical Systems recommends a Hypot Electrical Safety Checks be performed each time
the monitor’s case is opened or patient isolation is in question.
WARNING:
Do not touch the monitor when performing this test.
Disconnect all accessories from the monitor.
Plug the AC power cord from the CARDELL Monitor into the Hypot Tester.
WARNING:
Do not turn the monitor “ON” during a hypot test.
Perform a Hypot Test, per manufacturers’ instructions for one (1) minute, at 2.5 KV.
Verify that there was no arching observed.
HYPOT (SpO
2
)
WARNING:
Do not touch the monitor when performing this test.
Connect the appropriate SpO
2
probe and/or cable to the monitor.
Plug the AC power cord from the CARDELL Monitor into the Hypot Tester.
Connect a lead wire from the Ground terminal of the Hypot tester to the SpO
2
probe.
WARNING:
Do not turn the monitor “ON” during a hypot test.
Perform a Hypot Test, per manufacturer’s instructions for one (1) second, at 1.8 KV.
Verify that there was no arching observed.
Disconnect the monitor from the test equipment.
This concludes the testing to the CARDELL Monitor.