Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
66
SpO
2
AUDIO
(Model 9402)
Perform the following procedure to verify the audio range for the SpO
2
volume.
Press the VOLUME pushbutton twice. Verify a Key Click tone is heard each time and the
Message Window displays “SpO
2 ”.
Press the VOLUME UP and VOLUME DOWN pushbuttons and verify the SpO
2
“beep” volume
can be adjusted to one of five volume levels and “OFF”.
Set the volume level as desired.
Press the CANCEL pushbutton when completed.
CONFIGURATION MODE TESTS
Perform the following procedures to verify the following system operations of the monitor.
NOTE:
The monitor must be in the Configuration Mode in order to perform the following functions.
All LEDs “ON” Check
+12 Volt Power Supply Check
Calibration Check
System Pressure
Over Pressure
Pneumatic Pressure Checks
ENTERING THE TEST MODE
To enter the monitor Configuration Menu, press and hold the VOLUME and ALARM LIMITS
pushbutton keys while the monitor is being turned “ON”.
Once in the menu, press one of the PREVIOUS (CYCLE TIME / ALARM LIMITS) programmed
pushbutton keys until the Message Window briefly displays “Test Mode” followed by “0 mmHg”.
NOTE:
While in the Test Mode if no pushbutton is pressed within 15 minutes, the monitor will
automatically terminate the Monitor Configuration menu and return to the “Ready” mode.
WARNING:
Do not place the monitor in the TEST MODE when a cuff is attached to a patient.
Back
Next
Go To Table Of Contents