11. MAINTENANCE PROCEDURES
INTRODUCTION
This section discusses the tests used to verify performance following repairs or during routine
maintenance. All tests can be performed without removing the CARDELL Monitor’s cover.
If the CARDELL Monitor fails to perform as specified in any test, repairs must be made to
correct the problem before the monitor is returned to the user.
EQUIPMENT REQUIRED
To test the NIBP
P9 Calibration Kit
Mercury Manometer
NIBP Simulator
500 mL Pressure Cylinder (CAS p/n 01-02-0248)
To test the SpO
2
SpO
2
Finger Sensor Simulator
To perform Electrical Safety
Electrical Safety Analyzer
Hypot Tester
Data Sheet
This procedure uses a Data Sheet as the record for verifying monitor performance. Once the
procedure is completed, CAS recommends the Data Sheet be kept with the respective monitor’s
Device History Record should verification of monitor performance be questioned.
The DATA SHEET can be found on page 73.