ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)
ERROR
MESSAGE
POSSIBLE CAUSE
POSSIBLE SOLUTION
"RangeErr"
The systolic reading exceeds the
measurement range of 265 mmHg.
Repeat measurement.
If the message is displayed again, use another
method to measure the patient’s blood pressure.
“SetClock
The monitor’s clock needs to be
set.
The monitor’s time and date values are incorrect.
Refer to sections:
SETTING THE DATE and
SETTING THE TIME for more information.
The monitor’s internal clock battery needs to be
replaced. Contact CAS Medical Systems.
“Sig Sat”
Motion pulses too strong.
Limit patient activity; the arm must be still and/or
relaxed.
Repeat measurement.
“Sys Err”
An electronic failure has occurred
within the monitor.
Contact CAS Medical Systems to have the
monitor serviced.
"Time Out"
The monitor was unable to
complete a measurement within
150 seconds.
An extremely long measurement can be due to a
loose cuff, high blood pressure, or monitor re-
pumps.
Try measurement again. Try higher initial
pressure.
If message consistently reappears try using
another means to obtain patient's blood pressure.
"Weak Sig"
The monitor did not detect any
pulses during a NIBP
measurement.
Check the fit of the cuff.
Repeat measurement.
Table 5: Error Messages on the Message Window