Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
61
ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)
ERROR
MESSAGE
POSSIBLE CAUSE
POSSIBLE SOLUTION
"Motion"
There was too much extremity
motion for the monitor to
accurately complete the NIBP
measurement in 150 seconds.
Measurements can be obtained when there is
limited extremity movement, but the
measurement time may be extended.
Measurement time is limited to 150 seconds.
Restrain patient extremity motion.
"NBP Cal"
Pressure calibration data corrupted
within NIBP module.
Pressure module needs recalibration.
Contact CAS Medical Systems to have the
monitor serviced.
"NIBP Err"
An electronic failure has occurred
within the NIBP module.
Remove the battery and power cord from the
monitor. Reconnect the battery and cycle power.
Contact CAS Medical Systems to have the
monitor serviced.
“No Probe”
The monitor is not detecting the
SpO
2
probe.
The probe was disconnected from either the
Interface Cable or from the monitor.
"OverPres"
Cuff pressure exceeded 290
mmHg.
Very rapid squeezing of the cuff can cause this
error.
Repeat the measurement.
If this message repeatedly occurs during normal
use, the monitor must be serviced.
“P Search”
The monitor is searching for a
Pulse signal.
Normal at power-up as the monitor searches for
a pulse.
The probe position may have changed.
Check the probe site.
“Prb OFF”
The monitor is no longer receiving
a patient signal from the SpO
2
probe.
The probe is no longer in contact with the patient.
Check the probe site.
"Pwr Fail"
Power was disconnected from the
monitor.
Press the SILENCE/RESET pushbutton to clear
the message.
OR
Re-cycle the monitor’s power.
Table 5: Error Messages on the Message Window