PNEUMATIC TUBING
CAUTION:
If the hose should become grossly contaminated with blood or other bodily fluids, it should be
discarded.
NOTE:
CAS does not recommend submersion of the hose. Liquid should not be permitted to enter the
hose because instrument damage may occur. The hose should be allowed to thoroughly dry
before use.
Prior to each patient use, inspect the NIBP Inflation Hose for proper connection, cracks and
kinks. As necessary, clean the pneumatic tubing using a soft cloth dampened with a germicidal
solution.
PRINTER
When the printer becomes dirty, wipe with a soft dry cloth. For extreme dirt buildup, soak a
cloth with mild detergent, wring well and wipe. Dry by wiping with a soft dry cloth.
CAUTION:
Before cleaning the printer, disconnect the AC adapter from the printer.
Do not use volatile chemicals such as thinner, benzene, etc.
Never wet the inside of the printer mechanism.
Refer to the printer User’s Manual for more information.
SpO
2
INTERCONNECT CABLE
Prior to each patient use, inspect the SpO
2
Interconnect cable for damage. As necessary clean
the cable using a soft cloth dampened with a germicidal solution.
SpO
2
SENSORS
(Reusable)
As necessary, the sensor may be surface cleaned by wiping it with a 70% isopropyl alcohol pad.
Allow the sensor to dry prior to placement on a patient.
CAUTION:
Do not soak or immerse the sensor or its cable in any liquid solution. Do not attempt to sterilize.
Refer to the Directions For Use pamphlet enclosed with each sensor for more information.