Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
48
THE DISPLAY
CAUTION:
Use care when cleaning the display. Scratches may occur.
Occasionally, as needed, clean the display window using a soft, lint-free cloth sprayed with an
alcohol free glass cleaner. Do not use either isopropyl alcohol or solvent to clean the display.
Use of these cleaners can cause damage to the display. The use of paper towels is not
recommended as it may scratch the surface.
CUFFS
Prior to each patient use, inspect the blood pressure cuff for damage.
NOTE:
Refer to product packaging for additional Cleaning and Disinfecting Instructions were applicable.
CAUTION:
If the cuff should become grossly contaminated with blood or other bodily fluids, it should be
discarded.
NOTE:
CAS does not recommend submersion of the cuff. Liquid should not be permitted to enter the
cuff bladder because instrument damage may occur. The cuff should be allowed to thoroughly
dry before use.
TUFF-CUFF
As necessary, clean the cuff using a soft cloth dampened with a 70% Isopropyl Alcohol
solution.
SAFE-CUFF
As necessary, clean the cuff using a soft cloth dampened with a soap, water-based
detergent or chlorinated disinfectant solution. Do not use alcohol.
SOFTCHECK
These cuffs are designed for single patient use, and are not to be reprocessed.
ULTRACHECK
As necessary, for normal cleaning with mild detergents / dilute bleach solution (1-2%), wipe
the cuff with the cleaning solution, rinse with water and dry.
Back
Next
Go To Table Of Contents