Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
43
SET THE SpO
2
ALARM DELAY
(Nellcor SpO
2
only)
The delay time until an alarm is generated for %SpO
2
and Pulse Rate can be configured to be
either zero (0) seconds “Delay 0S” or ten (10) seconds “Delay10S”. The default value used by
the CARDELL Monitor is ten (10) seconds.
Should it need to be changed, first enter the Configuration menu. Refer to Page 39, ENTERING
THE MONITOR CONFIGURATION MENU.
Once in the menu, use one of the NEXT programmed pushbutton keys until the Message
Window displays the monitor’s current setting for the SpO
2
Alarm Delay.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
SETTING THE DATE
The CARDELL Monitor’s Date value is set at the factory. Should it need to be changed, enter
the Monitor Configuration menu and use one of the NEXT programmed pushbutton keys until
the Message Window displays the monitor’s date using the following format: “DDMMMYY”.
Where DD = Day of the Month, MMM = Month of the Year (Jan, Feb, etc.) and YY = Last 2
digits of the year (2002 is displayed as 02). The flashing parameter indicates the parameter that
can be changed.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection. Press one of the
NEXT programmed pushbutton keys to advance to the next parameter to set within the Date
menu.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
SETTING THE TIME
The CARDELL Monitor’s Time value is set for Eastern Time and is set at the factory. Should it
need to be changed, enter the Monitor Configuration menu and use one of the NEXT
programmed pushbutton keys until the Message Window displays the monitor’s time using the
following format: “HH:MM”. Where HH = Hour of the Day (0 – 23) and MM = Minute of the Hour
(0 – 59). The flashing parameter indicated the parameter that can be changed.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection. Press one of the
NEXT programmed pushbutton keys to advance to the next parameter to set within the Time
menu.
NOTE:
Altering the Date and Time will affect the History readings, but not erase them.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.
Back
Next
Go To Table Of Contents