ALARM LIMITS OFF
When the monitor is configured to the Alarm Limit Off setting, all alarms associated with patient
alarms are “OFF”. This mode is useful for spot check applications or if the monitor is being
moved from patient to patient and the user may not want to be disturbed by any audible alarms.
The SILENCE visual indicator, located on the front panel of the monitor will flash at a rate of two
(2) seconds “ON” and two (2) seconds “OFF” and the message “Alrm Off” will be displayed on
the Message Window as a reminder when enabled.
During an alarm limit off period, if an equipment alarm occurs, except for Low Battery and Dead
Battery, the audio alarm remains silenced and only a visual indicator is provided.
MAP VALUE ENABLE / DISABLE
During a blood pressure reading, the user can elect to display or not to display the MAP value.
Should it need to be changed, first enter the Monitor Configuration menu. Refer to Page 39,
ENTERING THE MONITOR CONFIGURATION MENU.
Once in the menu, use one of the NEXT programmed pushbutton keys until the Message
Window displays the monitor’s current setting for the MAP value “MAP On” or “MAP Off”.
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.
NOTE:
When “MAP Off” is selected, MAP values are omitted from History Display and Printing as well.
All alarms associated with MAP values are also disabled.
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the
CANCEL pushbutton to exit to the “Ready” mode.