7. MONITOR CONFIGURATION
The Monitor Configuration section allows the user to configure the CARDELL Monitor. Once
entered, the user can:
• Review the monitor’s internal Software Revisions
• Set the Operating Language
• Make selections for Audio Alarms
• Choose to display the MAP value
• Set the SpO
2
Alarm Delay (Nellcor)
• Set the Date
• Set the Time
• Set Daylight Saving Time Option
• Perform System Checks (Refer to CONFIGURATION MODE TESTS)
All LEDs “ON” Check
+12 Volt Power Supply Check
Manometer Mode
Pneumatic Pressure Check
ENTERING THE MONITOR CONFIGURATION MENU
To enter the monitor’s Configuration Menu, press and hold the VOLUME
and ALARM
LIMITS
pushbutton keys while the monitor is being turned “ON” .
Once in the menu, use one of the NEXT (HISTORY / VOLUME) or PREVIOUS (CYCLE TIME /
ALARM LIMITS) programmed pushbutton keys to advance onto the next or go back to the
previous parameter in the Configuration Menu.
NOTE:
While in the Configuration Menu, if no pushbutton is pressed within 60 seconds, the monitor will
automatically save all changes made and exit the Monitor Configuration menu. The Message
Window will briefly display “Saving” and return to the “Ready” mode.
SAVING YOUR CHANGES
When you have completed configuring the monitor, press the CANCEL pushbutton to exit and
lock in your selection(s). The Message Window will briefly display “Saving” and return to the
“Ready” mode.