32
BPM Red colored LEDs indicate the Pulse Rate in BPM
(beats per minute).
Green colored LEDs provide a visual indication of the SpO
2
signal
strength in a bar graph form (Model 9402).
LARGE
CUFF
A yellow LED indicator used to inform the user that the NIBP is
operating with a Child size cuff or larger.
SMALL
CUFF
A yellow LED indicator used to inform the user that the NIBP is
operating with a Neonatal size cuff.
Message Window area used to display various messages that aid
the user in monitor operation.
A tri-colored visual indicator used to display the status of the
power source and battery condition.
The status of the LED is:
GREEN = Monitor is connected to an AC wall outlet.
ORANGE = In Use on Battery
RED = Battery Low or Dead Battery
A Yellow LED visual indicator used along with the
SILENCE/RESET pushbutton to display the status of the Audio
Alarm Silence feature.
The status of the LED is:
“ON” continuously = 2 Minute Audio Silence
Flash one second “ON”/one second “OFF” = Permanent
Audio Alarm Silence