Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
23
4. DECLARATION OF CONFORMITY
Manufacturers Declaration of Conformity
Electronic Emissions and Immunity
The Model 9401/9402 Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the Model 9401/9402 Monitor should assure it is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment
RF emissions – CISPR 11 Group 1 The Model 9401/9402 Monitor uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions – CISPR 11 Class B
The Model 9401/9402 Monitor is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions
Complies
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV differential
mode
+/- 2 kV common
mode
+/- 1 kV differential
mode
+/- 2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
(>95% dip in
U
T
) for 0.5 cycle.
40% U
T
(60% dip in
U
T
) for 5 cycles.
70% U
T
(30% dip in
U
T
) for 25 cycles.
< 5% U
T
(> 95% dip in
U
T
) for 5 seconds.
< 5% U
T
(>95% dip in
U
T
) for 0.5 cycle.
40% U
T
(60% dip in
U
T
) for 5 cycles.
70% U
T
(30% dip in
U
T
) for 25 cycles.
< 5% U
T
(> 95% dip in
U
T
) for 5 seconds.
Mains power quality should be
that of a typical commercial or
hospital environment. If user of
the Model 9401/9402 Monitor
requires continued operation
during power mains interruptions,
it is recommended that the Model
9401/9402 Monitor be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.
Back
Next
Go To Table Of Contents