Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
20
CAUTION:
ACCIDENTAL SPILLS – In the event that fluids are accidentally spilled on the monitor, take the
monitor out of operation and inspect for damage.
BATTERY POWER – If the monitor will not be used or not connected to AC line power for a
period over six (6) months, remove the battery.
ELECTRICAL SHOCK – To reduce the risk of electrical shock, do not remove the back cover.
Refer all servicing to qualified personnel.
ELECTROMAGNETIC COMPATIBILITY (EMC) – The equipment needs special precautions
regarding EMC. Be aware that strong electromagnetic fields may interfere with monitor
operation. Interference prevents the clear reception of signals by the monitor. If the hospital is
close to a strong transmitter such as TV, AM, or FM radio, police or fire stations, a HAM radio
operator, an airport, or cellular phone, their signals could be picked up as signals by the
monitor. If you feel interference is affecting the monitor, contact your CAS Medical Systems
representative to check the monitor in your environment.
ELECTROSURGERY – Measurements may be affected in the presence of strong
electromagnetic sources such as electro surgery equipment.
GROUNDING – Do not defeat the three-wire grounding feature of the power cord by means of
adaptors, plug modifications, or other methods. Do not use extension cords of any type. Do not
connect the monitor to an electrical outlet controlled by a wall switch or dimmer.
INTERFACING OTHER EQUIPMENT – Monitoring equipment must be interfaced with other
types of medical equipment by qualified biomedical engineering personnel. Be certain to
consult manufacturers’ specifications to maintain safe operation.
STACKING – Where monitor is used adjacent to or stacked with other equipment, the monitor
should be observed to verify normal operation in the configuration in which it will be used.