Service manual

CARDELL MONITOR SERVICE MANUAL
21-02-0286 REV. 02 05/09
110
OPERATING MODES
Characteristic Specification
Patient: Veterinary
NIBP: Manual, STAT or Automatic (at preset intervals)
History: Review of previous measurements
%SpO
2
: Continuous Monitoring
Temperature: Predictive or Continuous Monitoring
POWER
Characteristic Specification
Source: External line or internal battery
AC Power: 100 - 240 VAC, 50/60 Hz, 0.5A; Fuse Rating – T500mAL250V or
T1.25AL250V (two provided) Refer to monitor rear panel labeling for actual
fuse rating.
Battery: Nickel Metal Hydride (NiMH) battery pack (user removable)
Charge Time: 4 hours
Operation on battery: 100 NIBP readings when set in the 5-minute
Automatic Mode
Leakage Current: 100 microamp (maximum)
FEATURES
Characteristic Specification
Self Test: System self test is performed each time power is turned on.
Auto Zero: Zero pressure reference is automatically established after every reading.
Inflation: Initial inflation to 150 mmHg or user selectable. (80, 100, 120, 140, 150,
160, 180, 200). Subsequent inflation to approximately 30 mmHg greater
than previous Systolic pressure.
Deflation: Automatic
Max Measurement Time: Limited to 150 seconds
Back
Next
Go To Table Of Contents