PATIENT ALARMS
9401 9402
Patient Parameters
Low Limit High Limit Low Limit High Limit
SYS: 30 – 260 30 – 260 30 – 260 30 – 260
DIA: 20 – 215 20 – 215 20 – 215 20 – 215
MAP: 25 – 230 25 – 230 25 – 230 25 – 230
Pulse: 35 – 295 35 – 295 35 – 295 35 – 295
% SpO
2
: 70 – 95 80 – 99
SpO
2
Pulse: 25 – 295 25 – 295
NOTE:
Each alarm limit may also be selected “OFF” individually or as a whole.
Low Limits cannot be set above the associated High Limit.
High Limits cannot be set lower than the associated Low Limit.
CONTROL PANEL
Characteristic Specification
Display: LED display of measurement results, instructions, troubleshooting
messages and signal strength bar.
Parameters Displayed: Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure (MAP)
Pulse Rate
%SpO
2
SAFETY LIMITS
Characteristic Specification
Automatic Cuff Deflation: If cuff pressure exceeds 290 mmHg
If measurement time exceeds 150 seconds
If safety timer detects microprocessor failure