EQUIPMENT ALERT
Checkonlyonehandpieceatatime.Donotadjust
regulatorover65PSI.
Step8:Checkhandpieceairflow.
A) Remove handpiece from holder.
B) Depress foot control.
C) Regulate air pressure by adjusting the corresponding valve.
D) Repeat for each handpiece.
Note: If the air adjustment valve is fully open and pressure is too low, then increase main
air regulator pressure until the recommended pressure is reached.
AirAdjustment
Valve
Decrease
Increase
Adjustment
Knob
Unlock
Lock
ToAdjustRegulator...
A) Pull adjustment knob up to unlock.
B) Rotate adjustment knob clockwise to
increase pressure, counterclockwise to
decrease pressure.
C) Push adjustment knob down to lock.
Air Regulator Adjustment
Note
The air adjustment valve that corresponds to the ultrasonic scaler (optional)
must be OPEN in order to operate the scaler with the foot pedal.
003-1878-99