4-4  HOW TO AVOID MALFUNCTIONS 
4-4-1  PREPARATION OF THE  
  SPECIMEN 
When preparing specimens, be sure that a suitable 
embedding medium, fixation, dehydration and 
infiltration time are chosen. 
4-4-2  TEMPERATURE OF THE  
  SPECIMEN 
Sectioning is carried out at ambient temperature 
(excluding frozen sections). If the temperature is 
too high, the paraffin softens. Therefore, avoid 
heating paraffin specimens by direct exposure to 
sunlight or other sources of heat. 
4-4-3  TIGHTENING THE 
  CLAMPING SCREWS 
Tighten all clamping screws and clamping levers on 
the knife carrier as well as specimen holder and 
specimen orientation. 
4-4-4  SELECTION OF THE KNIFE 
Carefully select the required knife material and 
profile. 
4-4-5  ADJUSTMENT OF THE KNIFE 
Take care to adjust the proper clearance angle of 
the knife: 5  - 7° for glass knives. For diamond 
knives ask the knife manufacturer. 
Select a clearance angle adjustment of 
5 - 15° according to the facet angle. 
Take care to adjust knife height. 
4-4-6  CUTTING SPEED 
Take care to select proper cutting speed. 
General Rule: The harder the material, the slower 
the cutting speed!! 
4-4-7  TRIMMING 
Take care in bringing knife and specimen together. 