Instruction manual

Rotary Microtome HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
CAUTION! The permanent functionality of the
automatic approach system is only guaranteed
when the working process is not disturbed by
section waste. Therefore, keep the automatic
approach system itself as well as the direct
surrounding area clean.
If the button (fig. 5.9) is pressed, although the
automatic approach system is not installed or not
yet ready for work, the display (fig. 5.7) shows NO
APPROACH. Again press button (fig. 6.9) to
delete this information. Then continue working as
usual without the automatic approach system.
CAUTION! Loosen the connecting cable on the
right side of the instrument, before removing knife
carrier or automatic approach system. For this, pull
off the grooved grip of the connector directly on
the housing to loosen the connecting lock.