Instruction manual

    Rotary Microtome HM 355 S 
MICROM International GmbH 
Robert-Bosch-Str. 49 
D- 69190 Walldorf 
2-11-4 READJUSTING SPECIMEN 
  CLAMPS 
Frequent use of the clamping lever (fig. 12.1) can 
lead to the fact that the specimen clamps cannot be 
clamped optimally anymore. If the necessary 
readjustments are not carried out, it might even be 
possible that the specimen clamp cannot be 
clamped anymore by means of the clamping lever 
(fig. 12.1). 
The clamping lever (fig. 12.1) should be in an 
almost upright position. To determine the clamping 
position of the clamping lever, adjust the inner 
screw (fig. 13.1) on the rear side of the orienting 
adapter by means of an allen key 
(size 3 mm). 
Turn the allen key in a clockwise direction if no 
clamping was achieved at all or if the clamping 
position of the clamping lever is too high. 
If the clamping position of the clamping lever is too 
low, turn the allen key in a counter-clockwise 
direction. 
  Fig. 13 
Then insert the specimen clamp and check 
clamping position. If necessary, repeat above-
described process. 