User Manual
33
Probe compensation
2.11 Probe compensation
Before connecting to any channels, users should make a probe compensation to insure the
probe match the input channel. The probe without compensation will lead to larger errors
or mistakes, and probe compensation optimizes the signal path and makes measurement
more precise. If the temperature is 10 degree centigrade or above, this progress must be
done to insure the accuracy of the measurement. See picture 2-12
Picture 2-12 Probe compensation menu
Procedures of probe compensation as follow
1) Connect the oscilloscope probe to Ch1. If you are using a hook head, users should
ensure the good connection with the probe.