User Manual
134
LIN bus trigger and decode
trigger. The trigger relation, trigger data configuration and data, trigger simultaneously.
e) 【1: data】—the measured signal data bits is nine. Only when the ninth bit is one, then
trigger. The trigger relation, trigger data configuration and data, trigger simultaneously.
f) 【x: data】— the measured signal data bits is nine. Only when the ninth bits is X, then
trigger. The trigger relation, trigger data configuration and data, trigger simultaneously;
g) Parity error — parity bit is valid at parity error, trigger while parity error.
The measured signal data bits is 8bit, parity bit, odd, baud rate, 19.2kb/s, trigger mode as
start bit, follow the steps as below
(1) In working mode, press , then press to select the trigger type
as serial bus.
(2) Press to select bus type as UART.
(3) Press to popup bus configuration menu, select Ch1, idle high,